PRP & PRF Tube Compatibility: TD4C Rotor Fit, Radius & RCF

 

VivaPRP technical reference

Most tube returns are not tube failures. They are one of two mismatches: a tube that does not seat in the rotor, or a protocol that asks for more force than the machine can deliver at that rotor.

This page covers both for the TD4C desktop centrifuge, using the rotor data published in its technical documentation. Educational content only, not medical advice. Confirm settings against your own validated protocol and the documentation shipped with your lot.

The two questions this page answers

  1. Fit — does the tube seat in the rotor? Usually decided by length, not diameter.
  2. Force — does the machine apply the × g your protocol was written for? Usually decided by rotor radius, not by the number on the display.

1. Fit: diameter is uniform, length is the variable

Every collection tube in our range shares a 16 mm outer diameter. What changes between sizes is length, and length is what the rotor seat constrains.

Configuration Volume Tube dimensions Seats in
ACD+Gel 8 ml 16 × 100 mm 8 × 10 ml / 8 × 15 ml / 8 × 20 ml
ACD+Gel 10 ml 16 × 100 mm 8 × 10 ml / 8 × 15 ml / 8 × 20 ml
ACD+Gel 12 ml 16 × 120 mm 8 × 15 ml / 8 × 20 ml
ACD+Gel 15 ml 16 × 120 mm 8 × 15 ml / 8 × 20 ml
Sodium Heparin+Gel 8 ml 16 × 100 mm 8 × 10 ml / 8 × 15 ml / 8 × 20 ml
Sodium Heparin+Gel 10 ml 16 × 100 mm 8 × 10 ml / 8 × 15 ml / 8 × 20 ml
GFC 10 ml 16 × 100 mm 8 × 10 ml / 8 × 15 ml / 8 × 20 ml
Activator gel 10 ml 16 × 100 mm 8 × 10 ml / 8 × 15 ml / 8 × 20 ml
PRF (additive-free) 10 ml 16 × 100 mm 8 × 10 ml / 8 × 15 ml / 8 × 20 ml
PRF (additive-free) 12 ml 16 × 120 mm 8 × 15 ml / 8 × 20 ml
PRF (additive-free) 15 ml 16 × 120 mm 8 × 15 ml / 8 × 20 ml

PRF tubes fit the TD4C

PRF is a separate product family, not a PRP variant: the tubes carry no anticoagulant and are spun immediately after draw. Physically they share the same 16 mm envelope, so they seat in the same 8-position rotors as the PRP tubes. Fit is not the variable with PRF — the protocol is. Section 4 covers how to set that protocol on this machine.

Read the rotor rating carefully

A rotor described as 8 × 20 ml holds eight positions, each rated to 20 ml — not twenty tubes. The position count is your batch size; the volume figure is headroom, not capacity.

The 12-position rotors (12 × 5 ml and 12 × 7 ml) are built for narrow-bore vacuum tubes. Standard 16 mm collection tubes do not seat in them without an adapter, and an adapter changes the geometry — see section 6.

2. The TD4C rotor table

Every TD4C rotor turns at the same maximum speed. They do not deliver the same force.

Rotor Positions × volume Max speed Max RCF (published) Implied radius
NO35701 8 × 15 ml 4,000 rpm 1,800 × g 10.1 cm
NO35702 8 × 20 ml 4,000 rpm 1,800 × g 10.1 cm
NO35703 8 × 10 ml vacuum 4,000 rpm 1,800 × g 10.1 cm
NO35703 12 × 5 ml vacuum 4,000 rpm 1,600 × g 8.9 cm
NO35703 12 × 7 ml vacuum 4,000 rpm 1,720 × g 9.6 cm
4 × 50 ml 4 × 50 ml 4,000 rpm Refer to rotor sheet —

Four thousand revolutions per minute, three different maximum forces: 1,600, 1,720 and 1,800 × g. The speed is identical in every row. What differs is the distance from the spindle to the tube bore — the narrow 5 ml and 7 ml positions sit closer to the centre of rotation, travel a shorter circle, and receive less force.

The 4 × 50 ml rotor is not covered by the published rotor table used above. Request its rotor sheet for the applied force before specifying a protocol for it.

Where the implied radius comes from

The implied radius is not a declared figure. It is what the published speed and force pair imply, solved from the conversion formula:

r = RCF ÷ (1.118 × 10-5 × rpm2)

Inserting the 8-position rotor’s 1,800 × g at 4,000 rpm returns 10.06 cm. The 12 × 5 ml rotor’s 1,600 × g returns 8.94 cm. The 12 × 7 ml rotor’s 1,720 × g returns 9.61 cm. Feeding each radius back into the formula reproduces the published figure, which is the cross-check that the three values are internally consistent. Treat the implied radius as a cross-check and the rotor documentation as the primary source.

3. rpm is not a portable setting

The same displayed speed delivers different force depending on the radius of the rotor it runs in.

Speed r = 8 cm (compact) r = 10.1 cm (TD4C, 8-position) r = 12 cm r = 16 cm (swing-out)
1,000 rpm 89 × g 112 × g 134 × g 179 × g
1,500 rpm 201 × g 253 × g 302 × g 402 × g
2,000 rpm 358 × g 450 × g 537 × g 716 × g
3,000 rpm 805 × g 1,012 × g 1,207 × g 1,610 × g
4,000 rpm 1,431 × g 1,800 × g 2,147 × g 2,862 × g

Read across the 1,500 rpm row: 201 × g on a compact 8 cm rotor, 253 × g on the TD4C’s 8-position rotor, 402 × g on a 16 cm swing-out. Same setting, double the force end to end.

This is why a protocol quoted only in rpm is under-specified, and why copying it onto a different machine is the most common reason a preparation separates differently than expected. For the formula, the full 1,000–4,000 rpm conversion table across five radii, and the reverse × g to rpm calculation, see the RPM to RCF converter.

4. Can your protocol run on this machine?

This is a capability check, not a settings table.

The ceiling comes first

The 8-position rotors top out at 1,800 × g; the 12-position rotors at 1,600 – 1,720 × g. If your protocol specifies a force above the ceiling of the rotor you own, this machine cannot produce it, whatever rpm you dial in. A larger number on the display does not create more force — only a rotor with a larger working radius does.

Set the machine in × g where your protocol is written in × g

The TD4C accepts a centrifugal-force setpoint as well as a speed setpoint. Where your protocol gives you a force, enter that force and let the machine apply its own rotor geometry — the conversion step, and the conversion error with it, disappears. Reserve rpm entry for protocols that were written on this machine in the first place.

Where the target number comes from

Published PRP and PRF protocols differ by tube format, anticoagulant system, gel formulation and the clinician’s own validated method. We do not publish a single spin setting per tube family, because one number applied across an entire range would be a marketing line rather than a specification. Your validated protocol supplies the target; the rotor table above tells you whether this machine can reach it.

PRF specifically

PRF tubes carry no anticoagulant and are spun immediately after draw, so the workflow differs from PRP even though the tube seats in the same rotor. Set the force from your PRF protocol, then either enter it directly in × g, or convert it to rpm using the 10.1 cm column above. Do not carry a PRP spin setting across to PRF — the two preparations are not interchangeable at the centrifuge.

What ships with the order

Lot documentation records the sterilisation cycle and dose for that production run, and the protocol card covers the validated conditions for the tube family supplied. If a card is missing from a delivery, contact our team before the first run rather than estimating the setting.

5. Five-point check before you order

  1. Measure the rotor seat, not the rotor. The constraint is the maximum tube length the seat accepts and the bore diameter, not the rotor’s outside dimension.
  2. Confirm the radius to the sample, from the rotor sheet or the machine’s × g display. On a swing-out rotor the tube swings level in operation, so the running geometry differs from the stopped one.
  3. Compare your protocol’s × g against the rotor ceiling before you commit to a rotor size. Above 1,800 × g, the 8-position rotors cannot help you.
  4. Check the fill line. An under-filled evacuated tube changes the blood-to-anticoagulant ratio, which affects platelet recovery independently of the spin.
  5. Run one box before you commit to volume. One box of each candidate size through your existing protocol answers the fit question more reliably than any table, including this one.

6. Common fit and force mistakes

Mistake What actually happens Fix
Quoting the machine’s headline maximum RCF for a smaller rotor You assume 1,800 × g and the narrow rotor only reaches 1,600 × g Read the ceiling for the rotor you actually own
Copying rpm from a protocol written for another centrifuge Force differs by up to 2× at the same displayed speed Convert to × g first, or enter the × g directly
Fitting 16 mm tubes into a 12 × 5 or 7 ml vacuum rotor The bore is narrow; an adapter is required and it shifts the sample radius Use an 8-position rotor for standard tubes
Reading “8 × 20 ml” as twenty tubes Batch size is eight positions, not twenty Plan runs around eight positions
Loading a rotor with empty seats Unbalanced rotor; separation quality drops or the run aborts Fill every position, or size the rotor to the batch
Re-spinning a gel tube to “improve” separation Breaks the barrier that already formed Fix the first spin parameters instead of adding a second

7. Frequently asked questions

Do PRF tubes fit the TD4C?

Yes. Our PRF tubes share the 16 mm outer diameter of the PRP tubes and seat in the same 8 × 15 ml and 8 × 20 ml rotors. What differs is the protocol, not the fit.

Will a 15 ml tube fit an 8 × 15 ml rotor?

Yes, subject to seat depth. The 15 ml tube measures 120 mm, and the rotor’s volume rating says nothing about how deep the seat is. Confirm on your own unit before a full-volume order.

What is 1,500 rpm in × g on the TD4C?

253 × g at the 8-position rotor’s 10.1 cm implied radius. On a compact 8 cm rotor the same setting gives 201 × g; on a 16 cm swing-out, 402 × g. Without a radius, “1,500 rpm” does not correspond to a single force.

Why do the TD4C rotors have different maximum RCF when they all run at 4,000 rpm?

Because the tube bore sits at a different distance from the spindle in each rotor. Narrow-bore 5 ml and 7 ml positions are closer to the centre, travel a shorter circle and receive less force — 1,600 and 1,720 × g against 1,800 × g for the wider 15 ml and 20 ml positions.

Why publish an implied radius rather than a declared one?

Because speed and force are the two figures a manufacturer declares; the radius is what those two imply. Use the implied value as a cross-check on our arithmetic and the rotor documentation as the primary source.

Can I run PRP and PRF tubes in the same batch?

They seat in the same rotor, but they follow different protocols — PRP carries an anticoagulant and PRF does not. Running them together means one of the two is outside its validated conditions.

8. Related reading